• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS INFINITY LAB CORE LICENSE 3.X; CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE Back to Search Results
Catalog Number 09180443001
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas infinity software allows configuring the structure and content of the message to be sent to an external host.In this case, the result message has been configured in order to send the "low" and "high" values used to define the normal reference range type.The investigation found that after the upgraded cobas infinity version is not correctly sending via hca (host connectivity agent) the "high" value of the normal reference range type when it is defined with the operator.The "high" reference range value is sent as the "low" reference range value.The external lis (laboratory information system) gets reference ranges from cobas infinity to print the report for the patient, therefore this report is not correctly printed in this scenario.This behavior has been considered a software issue.Cobas infinity is not able to identify the "high" value of the reference range type when it is defined using the operators.
 
Event or Problem Description
There was an allegation of a software issue for the cobas infinity core software.After upgrading to the new version of infinity software, the customer's reference ranges for hdl and ldl tests are not correctly being sent to the customer's lis (laboratory information system).This issue causes incorrect information on the customer's reports.No patient was affected by the issue.The issue was detected in a test environment.The software version is 3.03.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INFINITY LAB CORE LICENSE 3.X
Common Device Name
CALCULATOR/DATA PROCESSING MODULE FOR CLINICAL USE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE INSTRUMENT CENTER AG TEGIMENTA
forrenstrasse
na
rotkreuz 6343
SZ   6343
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15213873
Report Number1823260-2022-02401
Device Sequence Number17322673
Product Code JQP
UDI-Device Identifier07613336177129
UDI-Public07613336177129
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number09180443001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2022
Initial Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-