| Model Number |
26280 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in a vein.A 14x90/9fr uni plus 75cm wallstent endoprosthesis was advanced for treatment.However, the stent strut was kinked and could not be deployed.The procedure was completed with another of the same device.There were no patient complications reported.The patient was stable after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The stent was returned unsheathed on the delivery system.The stent was deployed without issue and no issues were noted with the stent.A visual examination identified no damage or issues with the stent cups or stent holder of the returned device.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile examination identified no issues.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.The target lesion was located in a vein.A 14x90/9fr uni plus 75cm wallstent endoprosthesis was advanced for treatment.However, the stent strut was kinked and could not be deployed.The procedure was completed with another of the same device.There were no patient complications reported.The patient was stable after the procedure.
|
| |
|