MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 26280
Medical Device Problem Codes Positioning Failure (1158); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a vein.A 14x90/9fr uni plus 75cm wallstent endoprosthesis was advanced for treatment.However, the stent strut was kinked and could not be deployed.The procedure was completed with another of the same device.There were no patient complications reported.The patient was stable after the procedure.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The stent was returned unsheathed on the delivery system.The stent was deployed without issue and no issues were noted with the stent.A visual examination identified no damage or issues with the stent cups or stent holder of the returned device.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile examination identified no issues.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a vein.A 14x90/9fr uni plus 75cm wallstent endoprosthesis was advanced for treatment.However, the stent strut was kinked and could not be deployed.The procedure was completed with another of the same device.There were no patient complications reported.The patient was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15214348
Report Number2124215-2022-28723
Device Sequence Number17358146
Product Code MAF
UDI-Device Identifier08714729204015
UDI-Public08714729204015
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2023
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0028033572
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date08/11/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR