| Model Number |
DO250R |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with do250r tc tenotomy scissdelcvd1/serr130mm.According to the complaint description, the scissor broke while cutting soft tissue.This occurred during a total shoulder replacement surgery.An additional medical intervention was required.An x-ray was taken intraoperatively; there was a delay of 10-15 minutes.Additional information was not provided.The adverse event is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.If the device is returned in the future, an investigation will again be completed at that time.Based upon the investigation results a capa is not necessary.
|
| |
|