EXACTECH, INC. TRULIANT TIB IMP PSC INSERT SZ 3.5, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
TRULIANT TIB IMP PSC INSERT SZ 3.5, 13MM |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a right knee revision.Unstable knee and posterior poly wear were indicated.The 13psc was swapped to 17psc.Initial implant date is unknown.The patient was last known to be in stable condition following the event.The explanted implants are not available for return as the hospital is keeping them.
|
| |
|
Additional Manufacturer Narrative
|
|
Hospital is keeping the devices.Concomitants medical devices: 02-012-60-1440 - logic stem ext 14mm x 40mm 6692318, 02-020-11-0335 - truliant ps cem fem ps cem right sz 3.5 6574008, 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t 6576806, 200-07-32 - advanced patella 32mm 3 peg implant 6737274, 201-78-15 - holding pin mini sharp point 4 pk 6800956, 201-78-81 - 3" trocar, mod.Hex 2pk 6743637, 204-70-00 - tibial stem ext.Screw 6827100.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
There were no device breakages or surgical delays.Patient was initially implanted on (b)(6), 2021, and the revision was approximately 1 year 4 months post the initial implant procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
After review of additional information received the following sections, have been updated accordingly: b5, h6, h7, h9.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d6a (remove erroneous implant date from initial report) h6: medical device problem code, component code, investigation conclusion the revision reported was likely the result of joint instability, which allowed for posterior wear of the tibial insert.However, this cannot be confirmed because the revised component was not returned to exactech for evaluation and no images or radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|