• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 14 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 14 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 07/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Report source: foreign - australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00206.
 
Event or Problem Description
It was reported that the patient underwent a right knee revision approximately 1 year post implantation due to instability.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: normal alignment without periprosthetic lucency or fractures.A definitive root cause cannot be determined.Per package insert, instability is known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 14 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15216372
Report Number3007963827-2022-00205
Device Sequence Number6329386
Product Code MBH
UDI-Device Identifier00889024468535
UDI-Public(01)00889024468535(17)251116(10)64843179
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2025
Device Model NumberN/A
Device Catalogue Number42522100914
Device Lot Number64843179
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight92 KG
-
-