Model Number 1506-00-006
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 07/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient was revised to address pain, instability and loosening of the tibial tray at the cement to implant interface.Unknown cement was used.Full revision performed leaving patella implant in place.Doi: (b)(6) 2014.Dor: (b)(6) 2022.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary, no device associated with this report was received for examination.All available x-rays and photographs were reviewed, and there is not evidence that the implant loosening is product related.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot, the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed this medwatch report was initially submitted on 9/23/2022.This report is being resubmitted as this report failed submission as a result of the electronic gateway.Please reference fda ticket csh-28330 for additional details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 6 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15217089
Report Number1818910-2022-15726
Device Sequence Number12198843
Product Code JWH
UDI-Device Identifier10603295042402
UDI-Public10603295042402
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-00-006
Device Catalogue Number150600006
Device Lot Number7767170
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date12/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/18/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 6 CEM.; ATTUNE PS FB INSRT SZ 7 6MM.; ATTUNE PS FEM LT SZ 7 CEM.; BONE CEMENT (MFG. AND PRODUCT NAME UNKNOWN).; UNK ATTUNE KNEE PATELLA.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
Patient Weight150 KG
-
-