| Model Number |
4251652-02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: event 5: detailed inquiry description - customer reports a total of 5 incidence from multiple clinicians of sheered catheter.No injury reported.Two possible lot numbers provided: lot 21m27g8361; expiry date 11/01/2026; production date 11/27/2021; lot 21n29g8392; expiry date 12/01/2026; production date 12/29/2021.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Event 5: this report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Two photographs displaying packaging and batch information were provided.Based on the data from the investigation results, the root cause of the reported incident was unable to be determine.The actual defective device is a valuable tool in investigating the cause of this incident.A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|