BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7213
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 80% stenosed target lesion was located in mildly calcified mid left anterior descending artery.Two 2.50mm x 15mm nc emerge balloon catheters were used for dilatation.However, upon insertion, both devices were fractured prior to crossing the lesion.It was noted that the first device broke closer to the wire exit port and the other device broke closer to the hub.The devices were removed and no fragments left inside the patient's body.The procedure was completed with a 2.5x15 non-boston scientific stent.There were no patient complications reported.
 
Event or Problem Description
It was reported that shaft break occurred.The 80% stenosed target lesion was located in mildly calcified mid left anterior descending artery.Two 2.50mm x 15mm nc emerge balloon catheters were used for dilatation.However, upon insertion, both devices were fractured prior to crossing the lesion.It was noted that the first device broke closer to the wire exit port and the other device broke closer to the hub.The devices were removed and no fragments left inside the patient's body.The procedure was completed with a 2.5x15 non-boston scientific stent.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The device was visually and microscopically examined.At 30.8cm from the strain relief, the hypotube of the device was kinked.At 34cm from the strain relief, the hypotube was separated.The separated ends were ovaled in shape indicating the device was kinked prior to separation.At 2.3cm, and 3cm distal from the separation, the hypotube was kinked.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded, and the device had tip damage.Photo media depicted the sizing on the hub as well as a hypotube kink, and a hypotube separation just proximal to the kink.The kink and separation aligned with what was seen in analysis and confirmed the reported separation.Product analysis confirmed the reported break as there was separation to the device.The device also had unreported hypotube kinks and tip damage.Photo media aligned with the reported events and confirmed the reported separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15217702
Report Number2124215-2022-29926
Device Sequence Number12440201
Product Code LOX
UDI-Device Identifier08714729846505
UDI-Public08714729846505
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2024
Device Model Number7213
Device Catalogue Number7213
Device Lot Number0028922004
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR