Catalog Number 07729952160
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation is patient/consumer.The meter was requested to be returned for investigation.
 
Event or Problem Description
We received an allegation of a display issue with a coaguchek vantus meter.The customer stated that he meter sounds like it turns on, but cannot see anything on the display.Customer is unable to see the menu or anything on the screen because it is black.Customer changed batteries three times due to the issue, but issue persists.
 
Additional Manufacturer Narrative
The meter was received for investigation.Sections d9 and h3 were updated.The device was turned on and the display remained black.There was no backlight.The circuit board was tested for damage or contamination.The battery compartment and printed circuit board were contaminated by liquid (leaked battery) which damaged the flex cable to the display and corroded the contacts (handling error).This caused an issue with the backlight and made the display unreadable.The investigation determined the issue was caused by a customer maintenance issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK VANTUS
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15218201
Report Number1823260-2022-02409
Device Sequence Number17614039
Product Code GJS
UDI-Device Identifier00365702700499
UDI-Public00365702700499
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170960
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number07729952160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-