Event or Problem Description
|
|
Reportedly, a 12.6mm tmicl12.6 -16.00/1.5/068 implantable collamer lens was implanted into the patients left eye (od) on (b)(6) 2022.The lens was exchanged for a different model, length and power lens on an unknown date.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Lens work order search: no similar complaint type events reported for units within the same lot.Claim#:(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: lens was returned in liquid, in vial.Visual inspection found a haptic and optic torn.Claim #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B4: corrected to 05-jul-2022 in initial mdr.B5: corrected to: reportedly, a 12.6mm tmicl12.6 -16.00/1.5/068 implantable collamer lens was implanted into the patients left eye (os) on (b)(6) 2022.The lens was exchanged for a different model, length and power lens on an unknown date.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.G3: corrected to 05-jul-2022 in initial mdr claim#: (b)(4).
|
| |
|