U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-64
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, the cs100 intra-aortic balloon pump (iabp) units screen is not working properly.It is unknown under which circumstances the event occurred or if there was any patient involvement.
 
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event or Problem Description
It was reported that before use, the cs100 intra-aortic balloon pump (iabp) units screen is not working properly.There is no patient involvement.
 
Additional Manufacturer Narrative
Updated sections : b4,e1,e2,e3,g4,h2,h11.Corrected sections: b5,b6,b7,d10,g6, h6 (impact & clinical code).
 
Additional Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to reproduce the reported issue.Manipulations as well as a restart of the cs100 made but still the same problem.Loaner device sent to customer pending.The fse reported that a maintenance shutdown letter on cs100 consoles has been sent to all owner/user customers.Getinge is therefore no longer involved in this product range.There will be no intervention from getinge part.Not returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15218397
Report Number2249723-2022-02055
Device Sequence Number20286509
Product Code DSP
UDI-Device Identifier10607567107509
UDI-Public10607567107509
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K031636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3013-64
Device Catalogue Number0998-00-3013-64
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer10/11/2022
05/09/2023
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date10/21/2022
05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/25/2010
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; UNKNOWN.
Patient SexUnknown
-
-