• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR PS KNEE; EMPOWR PS KNEETM FEMUR, NONPOROUS, 6L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR PS KNEE; EMPOWR PS KNEETM FEMUR, NONPOROUS, 6L Back to Search Results
Model Number 242-01-106
Health Effect - Clinical Code Fall (1848)
Date of Event 07/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - patient fell and broke her tibia and femur.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as a fall.The previous surgery and the surgery detailed in this event occurred 2 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a ncmr#63686 associated with the main part #242-01-106, empowr ps kneetm femur, nonporous, 6l which documents that out of 12 parts lot, 7 parts were rejected due to gouges on cam, near condyles.Later, the rejected parts were reworked and retouched through spar.All other items in the lot were met with the design, fit and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to a fall.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Agent has clearly mentioned that "patient fell" and due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR PS KNEE
Common Device Name
EMPOWR PS KNEETM FEMUR, NONPOROUS, 6L
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15218435
Report Number1644408-2022-01056
Device Sequence Number5745827
Product Code JWH
UDI-Device Identifier00888912171762
UDI-Public(01)00888912171762
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number242-01-106
Device Catalogue Number242-01-106
Device Lot Number859T1311
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2022
Initial Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
346-12-755 LOT 442V1021; 351-03-105 LOT 022T1133
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight77 KG
-
-