| Model Number |
TRULIANT PS POR FEM PS POR RIGHT SZ 2.5 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitants: 02-022-35-2511 - truliant tib imp ps insert sz 2.5 11mm, 02-022-55-2525 - truliant por tib tray size 2.5f/2.5t, 200-07-29 - advanced patella 29m 3 peg implant.The products are not returning as the customer is retaining the devices.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that a 57 yo female patient, initially implanted on (b)(6) 2021, underwent a right knee revision approximately 11 months post the initial implant procedure for pain and a loose femur.She was revised with competitors¿ devices.The patient was last known to be in stable condition following the event.The products are not returning as the customer is retaining the devices.
|
| |
|
Additional Manufacturer Narrative
|
|
After review of additional information received the following sections, have been updated accordingly: h6: the revision reported was likely the result of an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening and pain.However, this cannot be confirmed as the devices were not available for evaluation and images/pre-revision radiographs were unable to be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, h6 clinical code, device problem code, component code, types of investigation, investigation findings, investigation conclusion codes.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, the patient underwent a right total knee replacement surgery.Approximately 11 months later, the patient complained of pain and was revised due to a loose femur.Because the patient has filed a short form, it appears that they are alleging prosthesis wear of an exactech polyethylene device.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b5, h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
|
| |
|