Model Number TRULIANT PS POR FEM PS POR RIGHT SZ 2.5
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: 02-022-35-2511 - truliant tib imp ps insert sz 2.5 11mm, 02-022-55-2525 - truliant por tib tray size 2.5f/2.5t, 200-07-29 - advanced patella 29m 3 peg implant.The products are not returning as the customer is retaining the devices.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported that a 57 yo female patient, initially implanted on (b)(6) 2021, underwent a right knee revision approximately 11 months post the initial implant procedure for pain and a loose femur.She was revised with competitors¿ devices.The patient was last known to be in stable condition following the event.The products are not returning as the customer is retaining the devices.
 
Additional Manufacturer Narrative
After review of additional information received the following sections, have been updated accordingly: h6: the revision reported was likely the result of an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening and pain.However, this cannot be confirmed as the devices were not available for evaluation and images/pre-revision radiographs were unable to be obtained.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, h6 clinical code, device problem code, component code, types of investigation, investigation findings, investigation conclusion codes.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Event or Problem Description
As reported, the patient underwent a right total knee replacement surgery.Approximately 11 months later, the patient complained of pain and was revised due to a loose femur.Because the patient has filed a short form, it appears that they are alleging prosthesis wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b5, h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT PS POR FEM PS POR RIGHT SZ 2.5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15219053
Report Number1038671-2022-00914
Device Sequence Number15112224
Product Code JWH
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K181794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRULIANT PS POR FEM PS POR RIGHT SZ 2.5
Device Catalogue Number02-020-12-0325
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer03/16/2023
03/18/2026
04/21/2026
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date03/21/2023
04/14/2026
05/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-