Model Number CV-6913
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular during prime, when the circulation was stopped after priming, a leakage of priming solution was noted around the cuvette.It is unknown if there was any effect on the patient or results of the surgery.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.Product was changed out.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) e1 (initial reporter - added city) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to correction, additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 4210, 4307).Type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #3: 3331 - analysis of production records investigation findings: 4210 - leakage/seal investigation conclusions: 4307 - cause traced to component failure.The returned sample was visually inspected upon receipt with no anomalies noted.The unit was pressurized to 15.0 psi +/- 1, with air and submerged in a water bath for 30 seconds to detect any possible leaks.No leaks were noted.The h/sat probe was then attached to the unit and pressurized again to 15.0 psi with air and submerged in a water bath for 30 seconds.The pressure began to drop and a leak was noted near the loop portion of the unit.A retention sample from the same product code and lot number was evaluated using the same test method; no visual damage and no leak when pressurized with air and submerged in a water bath was observed.H/s cuvettes are 100% leak tested and visually inspected in process to ensure each part meets tcvs specifications.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on august 11, 2022.Upon further investigation of the reported event, the following information is new and/or changed: (device availability - added date returned to manufacturer).(date received by manufacturer).(indication that this is a follow-up report).(follow-up due to additional information).(device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
H/S CUVETTE 3/8X3/8-JAPAN
Common Device Name
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key15220658
Report Number1124841-2022-00139
Device Sequence Number6687910
Product Code DRY
UDI-Device Identifier00699753270121
UDI-Public(01)00699753270121
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K915265
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberCV-6913
Device Catalogue NumberN/A
Device Lot NumberZC15
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer08/11/2022
09/23/2022
Initial Report FDA Received Date08/12/2022
Supplement Report FDA Received Date08/30/2022
09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-