| Model Number |
LOGIC TIBIA IMPLANT PSC INSERT, SZ 2, 13MM |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No device return expected due to this being a legal case.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported via a legal notification, that a patient who had underwent a partial knee revision on (b)(6) 2018 in which an optetrak logic psc tibial insert was implanted, underwent a second revision surgery on (b)(6) 2022, approximately 3 years 5 months post the 2018 revision.Upon information and belief, plaintiff required revision surgery for issues including but not limited to polyethylene wear, bone loss, osteolysis, instability and/or component loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined, insufficient information.
|
| |
|
Event or Problem Description
|
|
This report is for the second revision reported.The first revision indicated was reported under 1038671-2022-00915.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation results - the revision reported was likely the result of femoral loosening and polyethylene wear as stated in the operative notes.The femoral loosening was likely the result of an insufficient bond between the femoral component and the bone and/or patient-related conditions.The extent and root cause of the polyethylene wear cannot be determined as the devices were not available for evaluation and images of the explanted devices were not provided.These devices are used for treatments, not diagnosis.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, e1, e2, e3, h6 - health effect - clinical code, medical device problem code, component code, type of investigation, investigation findings and h11.The reported event was unable to be confirmed due to limited information received from the customer.A definitive root cause was unable to be determined; however, may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of users and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|