• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number V3.0
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that an ophthalmic console switching off on its own.Timing of the procedure and patient impact were not reported.Additional information has been requested none received till date.Additional information has been received indicating that the procedure was completed by restarting the console.There was no patient harm.
 
Additional Manufacturer Narrative
The company service representative, examined the system.And confirmed and replicated the reported event.To resolve the issue, the nonconforming central processing unit (cpu) host was replaced.The system was then tested and met all product specifications.A system non-conformance-based review of the batch/lot/serial number was performed and did not reveal, any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed, prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar.The root cause of the reported event is attributed to a nonconforming central processing unit (cpu) host.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The company representative was able to replicate the reported event.The central processing unit (cpu) host was replaced to address the issue.The cpu host was returned for testing on this investigation.The system was tested and found to meet product specifications.Non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The cpu host was received for testing on this investigation.A visual assessment of the returned sample revealed debris.The battery voltage was measured, and the sample was installed into a calibrated console and turned on.The sample was tested through a 4-hour burn-in and a surgical test 3 times without issue.The sample was found to function as intended.The root cause of the reported event is attributed to internal wear/reliability issues with the system.How or when these wear/reliability issues occurred remains inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15222533
Report Number2028159-2022-01177
Device Sequence Number6721010
Product Code HQC
UDI-Device Identifier00380657521609
UDI-Public00380657521609
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K120912
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV3.0
Device Catalogue Number8065752160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2022
Supplement Date Received by Manufacturer10/24/2022
04/14/2023
Initial Report FDA Received Date08/13/2022
Supplement Report FDA Received Date11/18/2022
05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2014
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-