ALCON LABORATORIES IRELAND LTD. CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number CCWET0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Date of Event 06/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that, following intraocular lens implant (iol) procedure, the patient was having blurred vision.The iol was exchanged for another lens approximately 2 months following the initial implant procedure.The clinical reason given for the explant was ¿anisometropia¿.Additional information was requested.
 
Additional Manufacturer Narrative
Additional information provided in a.2., a.3., b.2., b.5., b.6., b.7.D.10.E.4.And h.10.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was reported that, reported that a [23.5] lens was exchanged for a [22.5] lens due to blurred vision, anisometropia.This does not meet criteria for reporting based on applicable medical device regulations.A lens exchange from one power to a different power is an attempt to adjust or fine tune the refractive outcome trying to achieve a desired target when the initial lens calculations did not achieve the intended outcome.There is no reported defect or fault with the lens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15222642
Report Number9612169-2022-00393
Device Sequence Number6687979
Product Code POE
UDI-Device Identifier00380652405843
UDI-Public00380652405843
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/05/2025
Device Model NumberCCWET0
Device Catalogue NumberCCWET0.235
Device Lot Number25307341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date08/13/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC; MONARCH III IOL CARTRIDGE D; MONARCH IV IOL INJECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-