PENUMBRA, INC. POD; HCG, KRD Back to Search Results
Catalog Number RBYPOD10
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using a pod coil and a lantern delivery microcatheter (lantern).During the procedure, while advancing the pod coil through the lantern to the target vessel, the physician experienced resistance.Subsequently, while retracting the pusher assembly of the pod coil through the lantern, the physician also experienced resistance and the coil unintentionally detached.Therefore, the lantern containing the detached pod coil was removed from the patient.The procedure was completed using another pod coil and another lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Additional Manufacturer Narrative
Evaluation of the returned pod coil confirmed that the embolization coil was detached inside the returned lantern.Further evaluation revealed that the pusher assembly ddt was fractured inside the lantern.If the pod coil is retracted against resistance, the pusher assembly ddt may fracture, and the embolization coil may detach.The root cause of the resistance was due to the ovalization on the returned lantern.Further evaluation revealed kinked throughout the length of the pusher assembly.This damage was likely incidental to the reported complaint.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15222667
Report Number3005168196-2022-00373
Device Sequence Number15125486
Product Code HCG
UDI-Device Identifier00814548019659
UDI-Public814548019659
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPOD10
Device Lot NumberF113998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer08/15/2022
Initial Report FDA Received Date08/13/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-