• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT315
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that following a cataract extraction with an intraocular lens (iol) implant procedure, the patient experienced light spreads sharply when looking at the light at night.Even after 5 months postoperatively, the patient continued to complain of the symptoms so that the iol exchange procedure was carried out.Additional information was received stating the patient symptoms resolved.
 
Additional Manufacturer Narrative
The lens was returned in a plastic specimen cup, wrapped in wet surgical gauzes.The lens was cut into pieces, typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.The product investigation could not identify a root cause for the reported complaint.Power and resolution testing could not be conducted due to the extensive optic damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provided that the toric (1.50cyl) 18.5 diopter, (1.5 extended depth of focus) lens was exchange for a non-company monofocal lens (19.0d).Due to the exchange of a toric extended depth of focus lens to a monofocal lens may have been an improved choice for the patient's vision needs.No additional information has been provided at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15222744
Report Number1119421-2022-01769
Device Sequence Number5745938
Product Code POE
UDI-Device Identifier00380652421614
UDI-Public00380652421614
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2026
Device Model NumberDFT315
Device Lot Number15195896
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/17/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date08/14/2022
Supplement Report FDA Received Date11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-