• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD NON-VENTED HOSPITAL FULL FACE MASK; CBK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD NON-VENTED HOSPITAL FULL FACE MASK; CBK Back to Search Results
Model Number RT041
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A distributor reported on behalf of healthcare facility in china that the seal of a rt041 full face mask disconnected from the mask frame.There was no reported patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint rt041 non-vented hospital full face mask was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and videograph provided by the customer, and our knowledge of the product.Results: visual inspection of the videograph revealed that the seal was partly separated from the mask base.It was further observed that a continuous bead of glue was found around the masks seals.Conclusion: we were unable to determine the root cause of the observed separation.Visual inspection indicated that the mask was assembled and glued properly.The rt041 full face mask features a mask base, seal, and headgear.The mask seal is inserted into the mask base and is secured with glue.Each assembled mask seal is inspected to ensure it is correctly inserted and there are no unacceptable gaps between the seal and the base.Samples are taken and the mask seals are pulled apart after 24 hours to ensure they meet or exceed the required detachment force before the assembled product can be released.Our user instructions illustrate in pictorial format the correct set-up and proper use of the rt041 non-vented hospital full face mask.It also states the following: "operating pressure range: 5 - 25 cmh2o." "this mask may only be used in a hospital or clinical setting where the patient is adequately monitored by trained medical staff.Failure to monitor the patient may result in loss of therapy, serious injury or death." "in the case of therapy device failure, the use of this mask requires the same level of attention and assistance as in the use of a tracheal tube.".
 
Event or Problem Description
A distributor reported on behalf of healthcare facility in china that the seal of a rt041 full face mask disconnected from the mask frame.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON-VENTED HOSPITAL FULL FACE MASK
Common Device Name
CBK
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15223043
Report Number9611451-2022-00752
Device Sequence Number11926099
Product Code CBK
UDI-Device Identifier09420012429865
UDI-Public(01)09420012429865(10)2101597034(11)210414
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083122
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT041
Device Catalogue NumberRT041
Device Lot Number2101597034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date08/14/2022
Supplement Report FDA Received Date08/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-