Model Number OPT844
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in china reported that the tubing of an opt844 optiflow adult nasal cannula was found damaged during use.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Product background: the opt844 optiflow adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow interfaces are designed for use with the airvo 2 series of humidification devices.Nhf therapy and the airvo 2 device are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The opt844 consists of a lightweight delivery tube which is connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's nares.Method: the complaint opt844 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer and our knowledge of the product.Results: visual inspection of the provided photograph revealed the opt844 was torn near the manifold.The customer reported that the clothing clip was not used.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.The reported event is most likely to have been caused by the tubing being subjected to excessive force during use.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject cannula would have met the specification at the time of production.The setup instructions in the user instructions which accompany the opt844 nasal cannula include the following steps: "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." the user instructions also contain the following warnings/cautions: "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW ADULT NASAL CANNULA
Common Device Name
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite #300
irvine, CA 92618
9494534002
MDR Report Key15223047
Report Number9611451-2022-00733
Device Sequence Number17339690
Product Code CAT
UDI-Device Identifier09420012432124
UDI-Public(01)09420012432124(10)2100778382(11)190522
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT844
Device Catalogue NumberOPT844
Device Lot Number2100778382
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Initial Report FDA Received Date08/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
F&P 900PT561 HEATED BREATHING TUBE AND CHAMBER KIT; F&P PT101 AIRVO 2 HUMIDIFIER
Patient SexUnknown
-
-