Model Number IPN056400
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
The blade did not make a 360 compliant cut on the iliac artery and at the same time the uncut flap got stuck in the mechanism, with the risk of a dissection of the arterial vessel necessary to carry out the graft of the kidney to be transplanted.The same event was reported on the contralateral artery.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.
 
Additional Manufacturer Narrative
(b)(4).The device history record of batch number 74f2100126 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.No corrective action can be established at this moment since the product sample is not available for evaluation.Product sample is necessary to perform a proper investigation to determinate the root cause and the corresponding corrective actions.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility neither is being manufactured at the time.
 
Event or Problem Description
The blade did not make a 360 compliant cut on the iliac artery and at the same time the uncut flap got stuck in the mechanism, with the risk of a dissection of the arterial vessel necessary to carry out the graft graft of the kidney to be transplanted.The same event was reported on the contralateral artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PU DP-52K DISP PUNCH 5.2MM
Common Device Name
INSTRUMENTS, SURGICAL, CARDIOV
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15223765
Report Number3004365956-2022-00056
Device Sequence Number14853651
Product Code DWS
UDI-Device Identifier14026704714114
UDI-Public14026704714114
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2026
Device Model NumberIPN056400
Device Catalogue NumberDP-52K
Device Lot Number74F2100126
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-