• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 303, INC. BD PYXIS¿ ANESTHESIA STATION ES; AUTOMATED DISPENSING CABINET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 303, INC. BD PYXIS¿ ANESTHESIA STATION ES; AUTOMATED DISPENSING CABINET Back to Search Results
Model Number PAS 4000
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Customer reported that a pyxis anesthesia 4000 system fingerprint reader, unexpected stopped responding during a scheduled hemicolectomy preventing user access to the station.Customer was able to retrieve the undisclosed required medication from a nearby station.The reported delay to patient care was 5 minutes.Patient or user harm was not reported.This event is recorded in complaint number (b)(4).A technical support specialist evaluated the device and observed that a nurse successfully logged into the station.The technical support specialist verified that there were no pending operating system updates as a precaution and rebooted the device.Device confirmed operational.
 
Event or Problem Description
Customer reported that a pyxis anesthesia 4000 system fingerprint reader, unexpected stopped responding during a scheduled hemicolectomy preventing user access to the station.Customer was able to retrieve the undisclosed required medication from a nearby station.The reported delay to patient care was 5 minutes.Patient or user harm was not reported.
 
Additional Manufacturer Narrative
The following fields have been updated with additional/ corrected information: a1, b5, d1, d4, d5, d9, e2, e3, g1, g6, h2, and h6.A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaints with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from 20-jul-2020 to 20-jul-2022 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that fingerprint reader stopped responding.A technical service specialist noticed that nurses were attempting to login to the station.Performed soft reboot into station application and manually rebooting to resolve the issue.The system functioned as intended after troubleshot by the technical service specialist.
 
Event or Problem Description
It was reported by the customer that bd pyxis¿ anesthesia station es fingerprint reader stopped responding unexpectedly during a scheduled hemicolectomy, which prevented user access to the station.The customer was able to retrieve the required medication from a nearby station.There were no adverse events or injuries reported based on this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PYXIS¿ ANESTHESIA STATION ES
Common Device Name
AUTOMATED DISPENSING CABINET
Manufacturer (Section D)
CAREFUSION 303, INC.
10020 pacific mesa blvd.
san diego CA 92130
Manufacturer (Section G)
CAREFUSION 303
10020 pacific mesa boulevard
san diego CA 92121
Manufacturer Contact
phillip emmert
10020 pacific mesa blvd.
san diego, CA 92130
MDR Report Key15223857
Report Number2016493-2022-99127
Device Sequence Number5566525
Product Code BRY
UDI-Device Identifier10885403477829
UDI-Public(01)10885403477829
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPAS 4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date03/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1