• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 6F X 105CM STRAIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 6F X 105CM STRAIGHT Back to Search Results
Model Number 190070
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital used endomyocardial biopsy forceps to sample myocardial tissue for patients on (b)(6) 2022.During the operation, the patient has been locally anesthetized, and after many times of triggering, the myocardial biopsy forceps are still unable to obtain tissue (the jaws cannot be opened).Experts on the stage repeatedly confirmed that it was a quality problem of myocardial biopsy forceps, and then exchanged a new myocardial biopsy forceps and successfully removed the tissue.
 
Additional Manufacturer Narrative
The sample is indicated as returned.As of the date of this report, the sample has not been evaluated.A follow-up report will be provided once the device has been reviewed.
 
Event or Problem Description
The hospital used endomyocardial biopsy forceps to sample myocardial tissue for patients on (b)(6) 2022.During the operation, the patient has been locally anesthetized, and after many times of triggering, the myocardial biopsy forceps are still unable to obtain tissue (the jaws cannot be opened).Experts on the stage repeatedly confirmed that it was a quality problem of myocardial biopsy forceps, and then exchanged a new myocardial biopsy forceps and successfully removed the tissue.
 
Additional Manufacturer Narrative
A review of the batch records confirmed that the product was manufactured according to specification and no deviations or anomalies were found.One opened device was returned for evaluation.Visual inspection found no damage to the product.Functionally, the device would either not open or open slowly, then suddenly snap open.A magnified view found that one of the pins in the clevis mechanism may have been over pressed, causing an intermittent drag when trying to actuate the device.The samples are being sent back to the supplier for further investigation.A scar# s-2020-00006 has been issued to notify the supplier viant as&o holdings, llc that an investigation is necessary.Per the supplier's investigation, they will take necessary corrective measures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWZ ENDOMYOCARDIAL BIOPSY FORCEPS 6F X 105CM STRAIGHT
Common Device Name
JAWZ ENDOMYOCARDIAL BIOPSY FORCEP
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key15224503
Report Number0001625425-2022-01059
Device Sequence Number5978081
Product Code DWZ
UDI-Device Identifier00886333206797
UDI-Public00886333206797
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K951447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number190070
Device Catalogue Number190070
Device Lot Number5052393
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-