• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW AM/AT PI MIDLINE 2-L: 5.5FR X 15CM; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW AM/AT PI MIDLINE 2-L: 5.5FR X 15CM; MIDLINE CATHETER Back to Search Results
Model Number IPN037698
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the midline catheter was placed by the user facility on (b)(6) 2022.The patient was discharged to a skilled nursing facility and returned to the hospital on (b)(6) 2022 because the white hub had fallen off in the skilled nursing facility.No patient harm was reported.The patient's condition was reported to be fine.
 
Additional Manufacturer Narrative
Qn#(b)(4).
 
Event or Problem Description
It was reported the midline catheter was placed by the user facility on (b)(6) 2022.The patient was discharged to a skilled nursing facility and returned to the hospital on (b)(6) 2022 because the white hub had fallen off in the skilled nursing facility.No patient harm was reported.The patient's condition was reported to be fine.
 
Additional Manufacturer Narrative
Qn# (b)(4).The customer returned one photo for analysis.The photo confirmed the proximal lumen was separated from the extension line.The customer also returned one 2-l cvc catheter for analysis.Signs of use in the form of biological material were observed in the extension lines.Visual inspection confirmed the proximal extension line was completely separated from the luer hub.The proximal luer hub was not returned.The total length of the catheter body measured 6.6875", which is within specifications of 6.625-6.875" per catheter product drawing.The outer diameter of the proximal lumen measured 0.09425", which is within specifications of 0.093-0.097" per product drawing.The inner diameter of the proximal lumen measured 0.056", which is within specifications of 0.055-0.059" per product drawing.This indicates that the wall thickness measured within specifications.The catheter was functionally tested per the instructions for use (ifu).The ifu provided with this kit states, "flush each lumen with sterile normal saline for injection to establish patency and prime lumen(s)." when the distal line was flushed no leaks or blocks were detected.A manual tug test confirmed the distal extension line was secure to its luer hub.A device history record review was performed with no relevant findings.The ifu provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause catheter component damage or breakage.If placement damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.Do not secure, staple, and/or suture directly to outside diameter of catheter body or extension lines to reduce risk of cutting or damaging the catheter or impeding catheter flow.Secure only at indicated stabilization locations." the report of an extension line/luer hub separation was confirmed by a complaint investigation.Visual analysis revealed that the proximal extension line was separated from the luer hub.The proximal luer hub was not returned for analysis.Without the luer hub returned, the root cause could not be determined.Teleflex will continue to monitor and trend complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AM/AT PI MIDLINE 2-L: 5.5FR X 15CM
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15224771
Report Number9680794-2022-00491
Device Sequence Number5977556
Product Code PND
UDI-Device Identifier50801902130162
UDI-Public50801902130162
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161313
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/31/2023
Device Model NumberIPN037698
Device Catalogue NumberCDC-41552-MPK1A
Device Lot Number13F22B0608
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-