| Model Number |
RUBELLA IGG |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a questionable rubella igg elecsys from cobas e 801 module serial number 1785-06.The result was 42 ui/l (reactive).The result by vidas method was 7 ui/ml (non-reactive).The result by abbott alinity was 4.7 ui/ml (non-reactive) the result by diasorin liaison was (non-reactive).The questionable result was not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Further testing of the patient sample found the platelia igg results were 9.04 and 9.99 iu/ml(negative), the mikrogen blot rubella igg result was reactive, and the neutralization assay showed evidence for anti-rubella igg antibodies.As the customer's positive result was reproducible, the result was deemed correct for the sample.As the customer's calibration and qc were within the acceptable range, there was no issue found with the reagent.D4 - expiration date was updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample was submitted for investigation and the elecsys rubella igg result was 35.6 iu/ml (reactive).The elecsys rubella igm result was 0.296 coi (non-reactive).The investigation is ongoing.
|
| |
|