| Model Number |
S702 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the battery of this implantable pacemaker was suspected to be depleting prematurely.The estimated battery longevity was noted to have decreased from 4.5 years to elective replacement indicator (eri) between device follow-ups.This pacemaker was successfully replaced and is expected to be returned to boston scientific for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that the battery of this implantable pacemaker was suspected to be depleting prematurely.The estimated battery longevity was noted to have decreased from 4.5 years to elective replacement indicator (eri) between device follow-ups.This pacemaker was successfully replaced and is expected to be returned to boston scientific for analysis.No additional adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|