Model Number TNT G5 STAT
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable elecsys troponin t stat assay results on a cobas e 411 analyzer, serial number (b)(4).The initial result was 56.08 pg/ml.The sample was repeated due to the laboratory's policy to repeat critical troponin t stat results.The repeated result was 92.28 pg/ml.The customer noticed the discrepancy and decided to repeat the sample again on a different analyzer.The repeated result, obtained on a different analyzer, was 49.74 pg/ml.This result was deemed correct.The erroneous results were not reported outside of the laboratory.
 
Additional Manufacturer Narrative
The field service representative replaced the reagent pack using a new lot and performed calibration, which appeared to resolve the issue.Qc, precision, and performance checks passed and the results were within specification.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TROPONIN T STAT ASSAY
Common Device Name
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15225922
Report Number1823260-2022-02439
Device Sequence Number5602584
Product Code MMI
UDI-Device Identifier04015630942695
UDI-Public04015630942695
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051752
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTNT G5 STAT
Device Catalogue Number07398000160
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Initial Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-