| Model Number |
85416 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On completion of the investigation a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
During a z-fen case the doctor tried to track the icast stent into the celiac artery when the stent detached prematurely from the balloon.The doctor then removed the catheter and the sheath with the stent in the sheath.The doctor then tried the same procedure again with a new icast stent and the same situation occurred again.The doctor then used a 7x38x120 icast stent that was successful.The patient did not have any complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the details and the correspondence the stents had dislodged inside the cook introducer sheaths while advancing through the described tortuous iliac artery during a fenestrated endograft procedure with the intended target being the celiac artery.The devices in question were not returned and no images of the devices in question were received.Being that this event occurred twice with one clinical procedure suggests that the sheath was in a very tight bend that may have contributed to the stent dislodgement.As the details provide a 7mm x 38mm icast was able to pass supports this suggestion.A review of the device history records going back to the stent sub assembly level shows that there were no non-conformances noted and the product met all quality and performance requirements.This includes a review of the stent crimping process as well as the advancement of the crimped stent through the labeled introducer sheath without sheath compatibility issues.A labeling review was conducted.The ifu was found to provide adequate instructions in regards to the proper handling and delivery of the stent and not to force passage or removal of the stent if resistance is met.A risk review found that the risk management documents for this product adequately address the reported defect and the severity.As the anticipated occurrence level has been exceeded there has been a corrective action request generated.Based on the information provided in the complaint and within the investigation performed, there is no evidence to conclude that the devices were faulty or manufactured improperly.Without the actual devices used in the case, images or the introducer sheaths the complaint cannot be confirmed.Based on the investigation the root cause is impossible to define.Based on the review, the product met all quality and performance requirements at the time of manufacture.H3 other text: device not returned.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|