| Catalog Number |
04882458001 |
| Medical Device Problem Codes |
Circuit Failure (1089); Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's meter was requested for an investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of blackened charging contacts on the base unit of the accu-chek inform ii meter docking station.It was reported that there were no signs of melted plastic on the base or any damage with the meters docked in the base.
|
| |
|
Additional Manufacturer Narrative
|
|
The base unit was returned for investigation.Visual examination of the base unit determined there were no signs of melting or burning.Corrosion was observed on the charging pins.Mishandling by the customer is the cause of the corrosion on the charging pins.
|
| |
|
Search Alerts/Recalls
|