| Model Number |
Coronavirus |
| Medical Device Problem Codes |
False Negative Result (1225); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: retain test of the #218542 kit resulted in expected results.Unable to duplicate customer complaint.Root cause for false negatives is unable to be determined through retain testing root cause: no problem found;unable to determine.Source: phone.
|
| |
|
Event or Problem Description
|
|
Reported false negative sars result.The customer communicated the result was positive by another molecular method.
|
| |
|
Search Alerts/Recalls
|