| Model Number |
TMICL13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted an implantable collamer lens into a patient's eye in (b)(6) 2022.The surgeon now reports a planned lens exchange for a shorter length lens.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5 - the reporter indicated that the surgeon implanted a 13.2mm tmicl13.2; -3.50/+1.00/077 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2022.The patient experienced excessive vaulting and elevated intraocular pressure.On (b)(6) 2022 the lens was exchanged with a shorter length lens and this resolved the problem.The cause of the event was reported as the device.Claim#: (b)(4).
|
| |
|