• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ YEAST ID; KIT, IDENTIFICATION, YEAST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ YEAST ID; KIT, IDENTIFICATION, YEAST Back to Search Results
Model Number 448316
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd phoenix¿ yeast id the customer received discrepant results between two instruments.Confirmatory tests determined that misidentification had occurred.This event occurred 1 time(s).There is no indication that false results were reported to health care practitioners.There is no indication of patient impact.The following information was provided by the initial reporter: "the customer use yeast id panel for yeast identification.The panel result from phoenix 100 is c.Krusei, and the result from phoenix m50 is p.Wickerhamii.The customer rechecked the result with maldi-tof, the final id is c.Krusei.".
 
Event or Problem Description
It was reported that while using bd phoenix¿ yeast id the customer received discrepant results between two instruments.Confirmatory tests determined that misidentification had occurred.This event occurred 1 time(s).There is no indication that false results were reported to health care practitioners.There is no indication of patient impact.The following information was provided by the initial reporter: "the customer use yeast id panel for yeast identification.The panel result from phoenix 100 is c.Krusei, and the result from phoenix m50 is p.Wickerhamii.The customer rechecked the result with maldi-tof, the final id is c.Krusei.".
 
Additional Manufacturer Narrative
H.6 investigation summary: this complaint is not confirmed.This complaint is for misidentification of c.Krusei as p.Wickerhamii when using phoenix panel yid (448316) batch number 1201556.The complaint batch was not available for investigation due to the batch being expired at the time of the investigation and is beyond our stability timeframe.As a result, this complaint is not confirmed for performance.A review of quality notifications revealed no quality notifications generated on the complaint batch.A review of complaints revealed one additional complaint on the complaint batch.Complaint trending was performed and no trends were identified associated with this defect.Bd id/ast plant quality will continue to monitor for trends and take action as necessary.Please continue to communicate any additional concerns.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ YEAST ID
Common Device Name
KIT, IDENTIFICATION, YEAST
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15229489
Report Number1119779-2022-01110
Device Sequence Number11289983
Product Code JXB
UDI-Device Identifier30382904483167
UDI-Public30382904483167
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2022
Device Model Number448316
Device Catalogue Number448316
Device Lot Number1201556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-