This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6, h10 evaluation of the returned products confirmed damage to the sterile packaging blister.Sterility has not been compromised, therefore this event is no longer considered reportable.Device history record was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the products when they left zimmer biomet was conforming to specifications.The root cause of the reported event is likely due to transit damage.A corrective action has been opened to further investigate this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|