Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during an angiogram of the left lower extremity via access in the right groin, a flexor ansel guiding sheath separated at the hub while the sheath was in the patient.Investigation evaluation: reviews of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation with the hub separated from the sheath.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ the ifu instructs the user ¿avoid applying traction to hub during removal.If resistance is anticipated or encountered during withdrawal of the flexor sheath, consider reinserting the dilator and removing the sheath and dilator as a unit.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the limited information provided and the results of the investigation, cook has concluded that a component failure unrelated to manufacturing contributed to this incident.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|