• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ YEAST ID; KIT, IDENTIFICATION, YEAST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ YEAST ID; KIT, IDENTIFICATION, YEAST Back to Search Results
Model Number 448316
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while testing with bd phoenix¿ yeast id there was a different yeast id result between phoenix 100 and phoenix m50.Confirmatory testing was performed using maldi-tof, the final id is c.Glabrata.There was no indication that results were reported to clinicians, and there was no report of patient impact.The following information was provided by the initial reporter: "the customer use yeast id panel for yeast identification.The panel result from phoenix 100 is c.Glabra, and the result from phoenix m50 is c.Firmetaria.".
 
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H.6 investigation summary: this complaint is not confirmed.This complaint is for misidentification of c.Glabrata as c.Firmetaria when using phoenix panel yid (448316) batch number 1236938.The customer stated that "the panel result from phoenix 100 is c.Glabrata, and the result from phoenix m50 is c.Firmetaria.The complaint batch was not available for investigation due to the batch being expired at the time of the investigation and is beyond our stability timeframe.As a result, this complaint is not confirmed for performance.A review of quality notifications revealed no quality notifications generated on the complaint batch.A review of complaints revealed no additional complaints on the complaint batch.Complaint trending was performed and no trends were identified associated with this defect.Bd id/ast plant quality will continue to monitor for trends and take action as necessary.H3 other text : see h.10.
 
Event or Problem Description
It was reported that while testing with bd phoenix¿ yeast id there was a different yeast id result between phoenix 100 and phoenix m50.Confirmatory testing was performed using maldi-tof, the final id is c.Glabrata.There was no indication that results were reported to clinicians, and there was no report of patient impact.The following information was provided by the initial reporter: "the customer use yeast id panel for yeast identification.The panel result from phoenix 100 is c.Glabra, and the result from phoenix m50 is c.Firmetaria.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ YEAST ID
Common Device Name
KIT, IDENTIFICATION, YEAST
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15229842
Report Number1119779-2022-01112
Device Sequence Number4899448
Product Code JXB
UDI-Device Identifier30382904483167
UDI-Public30382904483167
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2022
Device Model Number448316
Device Catalogue Number448316
Device Lot Number1236938
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date08/15/2022
Supplement Report FDA Received Date01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-