• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cartridge of the preloaded intraocular lens (iol) device was broken/cracked.The lens was partially inserted into patient's right eye, however, was removed and another lens of same model and diopter was used as the replacement.The patient has recovered.No other information was provided.
 
Additional Manufacturer Narrative
Weight and ethnicity: unknown/not provided.If implanted, give date: not applicable, as there is no indication that the lens was implanted if explanted, give date: not applicable, as there is no indication that the lens was implanted.Device evaluation: the complaint handpiece was received inside of the original folding carton.Visual inspection under magnification revealed that the handpiece was received with the plunger rod in the retracted position and with the inside of the cartridge coated in viscoelastic residue.The cartridge tip was observed to be cracked and deformation could be observed on the cartridge, in a way consistent with a handpiece that was used.No further issues were observed with the handpiece.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.The search revealed no additional investigation request (ir) for this production order number has been received.Conclusion: the reported complaint issue was not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15230406
Report Number3012236936-2022-02147
Device Sequence Number15028832
Product Code HQL
UDI-Device Identifier05050474635968
UDI-Public(01)05050474635968(17)240803
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2024
Device Model NumberDCB00
Device Catalogue NumberDCB0000140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2022
Initial Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-