Model Number DCB00
Device Problems Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Event Description
It was reported that the preloaded intraocular lens (iol) had trailing haptic missing when inserted into patient's operative eye.The haptic was still left in the device.There was no other resistance occurred during the deployment and customer thinks that plunger overrode the optic and snapped off the haptic at the junction.No other intervention was required and the lens remains implanted.No other information was provided.
 
Manufacturer Narrative
If explanted, give date: not applicable, remains implanted in the eye.The device was not returned for evaluation as the lens remain implanted in the eye.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15230604
MDR Text Key305062729
Report Number3012236936-2022-02176
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474636088
UDI-Public(01)05050474636088(17)241210
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-