COVIDIEN NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
|
Back to Search Results |
|
| Model Number |
AB14W060200150 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use a nanocross elite pta balloon along with a non-medtronic 6fr sheath during procedure to treat a little calcified plaque lesion in the left mid superficial femoral artery (sfa).The vessel was none tortuous.The vessel diameter and lesion length are 5.5mm and 250mm respectively.No embolic protection was used.A non-medtronic inflation device was used for balloon inflation.There was no damage noted to packaging.There was no issue noted when removing the device from the hoop/tray.The device was prepped per ifu with no issues noted when removing from hoop/tray.The device did not pass through a previously deployed stent.There was no resistance encountered when advancing the device.It was reported that balloon twist occurred with no rewrap issue noted.Moved balloon to ensure all of the vessel was ballooned.Once inflations were complete, balloon was removed and inflated on back table.At that time the balloon inflated completely.We watched it unwrap and inflate.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis a visual examination of the balloon was carried out and a twist was noted approximately 8.5cm from the balloon tip a 20ml water filled syringe was used to flush the device and a steady stream of water was observed exiting the tip a 0.014¿ guidewire was loaded, and an indeflator with pressure gauge was used to inflate the balloon to its nominal pressure of 8atms and the balloon inflated without issue and the twist disappeared, the balloon was then inflated to its rated burst pressure (rbp) of 14atms, and the balloon inflated without issue and no twist.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|