|
SMITH & NEPHEW, INC. JOURNEY FLEX-EXT SPACER 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENT, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
74018608 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during surgery, it was noticed that the journey flex-ext spacer 10mm, the journey flex-ext spacer 9mm and the gii mis tib blck align rod were worn due to wear.The procedure was finished with a smith and nephew back up devices without delay.The patient was not harmed.Initially it was reported that the journey flex-ext spacer 10mm and the journey flex-ext spacer 9mm were worn, but during investigation it was noticed that both devices are broken in half.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode a visual inspection confirmed the device is broken in half, rendering it inoperative.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|