It was reported that an erbe system, argon plasma coagulator (apc)/electrosurgical unit (esu/generator, model vio 300 d, part number (p/n) 10140-100, serial number (b)(4)) system was involved in a patient incident upon an esophagogastroduodenoscopy (egd) and colonoscopy (patient had a history of anemia.).An apc probe was used in the procedures.Argon plasma coagulation was applied to address gastric, small bowel, and colonic arteriovenous malformations (avms).The patient was discharged from the hospital the same day.The next day, she was admitted at another hospital ((b)(6) hospital) and treated for a delayed perforation.A resection procedure was performed to remedy a perforation in the small bowel.
|
There is a plan to have the erbe equipment checked/evaluated at erbe usa.If any issues are found, a follow up report will be provided.No anomalies were found in device history records (dhrs) of the involved devices.Most likely, there were many factors involved in the event.However, the patient's condition appears to have played a significant role in the incident.The patient is very old, has a history of anemia/indicating bleeding, has avms throughout gi tract (and especially in thin-walled areas), etc.Specifically, upon the intervention work, the remaining tissue of the small bowl did not stay intact which resulted in the perforation.Nevertheless, at this time, no determination could be made as to the cause of the incident.To further address the issue, additional in-service work is being planned at the account with the involved staff on 08/16/2022.No trends have been identified and erbe usa, inc.Is now closing the file on this event.
|