Model Number 96-6510
Medical Device Problem Codes Break (1069); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported instrument is worn due to general wear and tear and an attachment is broken.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was not returned for analysis.Review of the photographic evidence confirmed, the complaint.The slap hammer stop component, broken off from the slap hammer external rod.Overall appearance of the device denotes wear and heavy use.Depuy synthes considers, the investigation closed at this time.Should additional information be received, the information will be reviewed.And the investigation may be re-opened as necessary.Device history lot: a manufacturing records evaluation (mre) was not performed, as no lot number was provided for this device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was not returned for analysis.Review of the photographic evidence confirmed the complaint.The slap hammer stop component broken off from the slap hammer external rod.Overall appearance of the device denotes wear and heavy use.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SP2 SLAP HAMMER ATTACHMENT
Common Device Name
KNEE INSTRUMENT : IMPACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15232178
Report Number1818910-2022-15871
Device Sequence Number17335538
Product Code FZY
UDI-Device Identifier10603295248620
UDI-Public10603295248620
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-6510
Device Catalogue Number966510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer10/06/2022
10/31/2022
Initial Report FDA Received Date08/16/2022
Supplement Report FDA Received Date10/07/2022
10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-