| Model Number |
1217-01-050 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Unspecified Musculoskeletal problem (4535)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
70-year-old female patient received a left hip revision to treat impingement secondary to dislocation and liner disassociation.Upon entering the joint, the surgeon identified and debrided metallosis caused by metal debris.The liner was disassociated due to locking mechanism failure, causing the summit femoral stem to impinge on the failed locking ring generating wear on the trunnion and metal debris.The cup and stem were well-fixed and retained.The patient a bimentum dual mobility articulating surface to replace the revised head and liner.The procedure was completed without complications.Head, liner and stem were revised.Doi: (b)(6) 2008(cup, stem).Doi: (b)(6) 2017(head, liner).Dor: (b)(6) 2021.Left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|