Model Number DM3500
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for an explant procedure.During the procedure, the implantable cardiac monitor's (icm) header came off.The icm was explanted.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of detached header was confirmed.Visual inspection found the device with a broken header which is consistent with explant damage.Longevity assessment was performed and device displayed normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15232297
MDR Text Key302865615
Report Number2017865-2022-19441
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/16/2020
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000068982
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
-
-