ETHICON ENDO-SURGERY, LLC. ECHELON FLEX 45 COMPACT; STAPLE, IMPLANTABLE Back to Search Results
Model Number SC45A
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the sales rep that during the lap appy the staple locked after firing.It was not able to open.Unknown if the procedure was completed and if there was any patient harm.
 
Additional Manufacturer Narrative
(b)(4).Batch # unk.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was the device locked on tissue with the jaws closed? yes, surgeon was able to cut the tissue and remove the stapler.If yes, how was the device removed? - yes.What troubleshooting steps were attempted to open the device (e.G.Squeezing the closing trigger while simultaneously depressing the anvil release button, knife reverse switch, manual override)?- yes we tried all the regular troubleshooting.Did the jaws eventually open?-yes, outside the cavity.Was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.)- no negative outcomes.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX 45 COMPACT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15232739
Report Number3005075853-2022-05309
Device Sequence Number15029000
Product Code GDW
UDI-Device Identifier10705036002697
UDI-Public10705036002697
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSC45A
Device Catalogue NumberSC45A
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-