| Model Number |
VICM5_12.6 |
| Medical Device Problem Codes |
Off-Label Use (1494); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This lens model is contraindicated for patients with age more than that of 45 years.Work order search: no similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon inserted and removed a 12.6mm vicm5_12.6 implantable collamer lens, -12.00 diopter, within the same surgery due to the lens flipped in the eye (lens rotation not associated to a low vault).There was no patient injury.The lens was replaced with another same model/length lens and the problem was resolved.Cause of the event was due to the device.
|
| |
|