Model Number 150419
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a visiting nurse checked the actual product for a patient at home and confirmed that the urination port of the leg bag was distorted and could not be used.Per follow up information received on (b)(6) 2022, there was no impact on patient and looked like there was no bent and twisted from the photo samples.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.It is unknown if the product was used for treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (with original packaging), used dispoz-a-bag.Visual inspection of the sample noted the flip-flo valve bend/ distorted.The photo sample provided shows the flip-flo valve bend/ distorted.This does not meet specification which states "damages are not allowed in the components".A potential root cause for this failure mode could be ¿inappropriate package design¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "directions for use: 1.Separate notches within circles.Pull straps through holes and around leg.2.Position bag on leg with flutter valve at top.3.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard extension tubing (catalog no.150615 or 4a4194).4.To empty dispoz-a-bag, push green lever on flip-flo valve out and down.Important: be sure to reclose flip-flo valve after emptying bag.5.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable, local, state and federal laws and regulations.Bard dispoz-a-bag accessories: ¿ leg bag holder ¿ 1-3/4¿ wide leg bag strap ¿ deluxe fabric leg bag straps ¿ extension tubing" the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that a visiting nurse checked the actual product for a patient at home and confirmed that the urination port of the leg bag was distorted and could not be used.Per follow up information received on 08aug2022, there was no impact on patient and looked like there was no bent and twisted from the photo samples.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEG BAG WITH FLIP-FLO VALVE
Common Device Name
LEG BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15233982
Report Number1018233-2022-06437
Device Sequence Number4692967
Product Code FAQ
UDI-Device Identifier00801741080654
UDI-Public(01)00801741080654
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2026
Device Model Number150419
Device Catalogue Number150419
Device Lot NumberNGFS4410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date08/16/2022
Supplement Report FDA Received Date02/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-