MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-2-6-HX-ES
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Deformation (2976); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received report that once the package was opened, the axium coil pushwire was observed to be kinked.The coil was pushed out and found to be already detached.It was noted the device was prepared per ifu but was not used in the procedure.No patient was involved.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that there was no damage or evidence of tampering to device packaging.No prep performed prior to the damage being seen.The proximal end of the pushwire secured in the guidewire clip (black rubber stopper) when the package was opened.The pusher wire kinked.
 
Additional Manufacturer Narrative
H3.Product analysis:equipment used: vis (m-78210), 203cm ruler (m-83361) as found condition (condition of returned device): an axium prime coil was returned for analysis within a shipping box; within an opened axium prime coil outer carton; within an opened axium prime coil inner pouch and within a dispenser track.Visual inspection/damage location details: the axium prime coil was returned within introducer sheath.The pushwire was found to be kinked at ~1.1cm from proximal end.The actuator interface, positive load indicator, and coupler tube were found to be intact.This is indicative mechanical method detachment was not attempted.The axium prime coil pushwire was found to be broken but retained by release wire at load indicator.The coin was found to be still against the lumen stop.The implant coil was already detached and not returned.No other anomalies were observed.Testing/analysis (including sem reports): under the microscope, the outer jacket was then removed to gain access to the coin.The coin appeared to be damaged and lost teardrop shape.The coin was unable to be measured accurately.The lumen stop appeared to be visually acceptable.The retainer ring was found to be visually acceptable and measured to be 0.0052¿ which is within specification.Conclusion: based on the analysis performed, the customers report of ¿premature detachment¿ was confirmed as the axium prime was returned with the implant already detached; however, the cause could not be determined.Since the implant coil and included detach, element were not returned for analysis, any contributing factors could not be determined.Based on the analysis performed, the customers report of ¿product deformed/damaged out of pkg¿ was confirmed as the pushwire was found to be returned damaged (kinked/broken), and not secured within the rubber stopper; however, the cause could not be determined.Since the implant coil and included detach element were not returned for analysis, any contributing factors could not be determined.Based on the device analysis and reported information, the customer¿s report of ¿pusher kink¿ was confirmed.Pushwire damage can occur if the user advances the device against resistance.However, there was not any friction or difficulty during testing or delivery.Therefore, the cause for the pushwire kink could not be confirmed.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15234503
Report Number2029214-2022-01382
Device Sequence Number17339713
Product Code KRD
UDI-Device Identifier00847536021013
UDI-Public00847536021013
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/15/2024
Device Model NumberAPB-2-6-HX-ES
Device Catalogue NumberAPB-2-6-HX-ES
Device Lot Number222935132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer08/25/2022
01/11/2023
Initial Report FDA Received Date08/16/2022
Supplement Report FDA Received Date09/07/2022
02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-