Model Number H7493941812350
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary artery.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, the device failed to cross the lesion and upon removal, the hypotube section of the shaft was fractured.The physician grabbed the remaining wire to pull out the end with the stent.The procedure was completed with another of the same device.No patient complications were reported.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary artery.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, the device failed to cross the lesion and upon removal, the hypotube section of the shaft was fractured.The physician grabbed the remaining wire to pull out the end with the stent.The procedure was completed with another of the same device.No patient complications were reported.It was further reported that the break occurred approximately 129 cm from the hub.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd mr us 3.50 x 12mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage with proximal struts lifted.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft and a break 22.5 cm distal to distal strain relief.A visual and tactile examination of the outer and mid-shaft section found no issues.No other issues noted with device.B5.Describe event or problem- updated.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary artery.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, the device failed to cross the lesion and upon removal, the hypotube section of the shaft was fractured.The physician grabbed the remaining wire to pull out the end with the stent.The procedure was completed with another of the same device.No patient complications were reported.It was further reported that the break occurred approximately 129 cm from the hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15234731
Report Number2134265-2022-08503
Device Sequence Number6446744
Product Code NIQ
UDI-Device Identifier08714729981039
UDI-Public08714729981039
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2023
Device Model NumberH7493941812350
Device Lot Number0027715496
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer08/22/2022
09/21/2022
Initial Report FDA Received Date08/16/2022
Supplement Report FDA Received Date09/12/2022
10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-