• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202901
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during inspection for an arthroscopy procedure, it was identified that the primary package of the footprint suture anchor was ripped and the sterility of the product was affected.The procedure was completed without surgical delay using a back-up device.No further complications were reported.
 
Manufacturer Narrative
H10: d4 information added.
 
Manufacturer Narrative
H10: internal complaint reference:(b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review and device sterility record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Manufacturer Narrative
Internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the packaging assembly found that boxing associate must visually confirm the presence of all seals.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15234811
MDR Text Key298292075
Report Number1219602-2022-01170
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010650528
UDI-Public03596010650528
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K093897
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202901
Device Catalogue Number72202901
Device Lot Number2078963
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/22/2022
Initial Date FDA Received08/16/2022
Supplement Dates Manufacturer Received08/26/2022
09/16/2022
12/20/2022
10/26/2023
Supplement Dates FDA Received08/26/2022
09/21/2022
12/20/2022
10/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-